MATERIALS REQUIRED BUT NOT PROVIDED
1.
Clock or timer.
2.
Specimen collection containers.
WARNINGS AND PRECAUTIONS
1. For in vitro diagnostic and
professional use only.
2. Do not
use the test device beyond the expiration date.
3. Urine
specimens may be infectious; properly handle and dispose of all used
reaction devices in a biohazard container.
4.
Visually
inspect the foil package to insure it is intact. If the package is not intact, the integrity of the device
might be compromised.
Storage and stability
Store
test kit below 28°C;
do not freeze. Refer to the
expiration date for stability.
Test procedure
1. Bring a urine sample and a foil test pouch to
room temperature (15°-28°C).
2. Do not break the seal of the pouch until
ready to begin testing.
3. Remove a Test Device from the foil pouch.
4. Remove the protective cap and place the
revealed strips into the urine sample for 10 seconds. Do not allow the
urine level to touch the plastic device.
5. Read the results at 5 minutes.
Note: The
result must be interpreted at five minutes. Waiting more than five minutes
may cause the reading to be inaccurate. To avoid confusion, discard the
test device after interpreting the result.
INTERPRETATION
OF RESULTS
POSITIVE
NEGATIVE
INVALID
Positive: A rose-pink band is
visible in each control zone. No color band appearing in the appropriate
test zone indicates a positive result for the corresponding drug of that
specific test zone.
Negative: A rose-pink band is visible in each control zone and the appropriate
test zone, indicating that the concentration of the corresponding drug of
that specific test zone is below the detection limit of the test.
Invalid:
If a color band is not visible in each of the control zones, the
test is invalid. Another test should be run to re-evaluate the specimen.
Note: There is
no meaning attributed to line color intensity or width.
Quality control
An
internal procedure control has been incorporated into the test to ensure
proper kit performance and reliability.
The use of an external control is recommended to
verify proper kit performance. Quality control samples should be tested
according to quality control requirements established by the testing
laboratory.
Limitations of the test
1. This product is designed to be used for the detection of drugs
of abuse and their metabolites in human urine only.
2. Although the test is very accurate, there is the possibility
false results will occur due to the presence of interfering substances in
the specimen sample.
3. The test is a qualitative screening assay and is not suggested
for quantitative determination of drug levels in urine, or the level of
intoxication.
4. Adulterants such as bleach or other strong oxidizing agents,
when added to urine specimens can cause erroneous test results regardless
of the analysis method used. If
adulteration is suspected, obtain another urine specimen.
Performance characteristics
1.
Sensitivity. The Multi-drug Test detects drugs of abuse and their major
metabolites in urine at concentrations equal to or greater than the
cut-off level for the specific drug, which is suggested by the National
Institute on Drug Abuse (NIDA) for the immunoassay method.
2.
Specificity. A study was conducted with the Multi-drug Test to determine
the cross-reactivity of drug-related compounds with the test.
Substances listed in Table I produced results approximately
equivalent to the cutoff levels. A separate study was conducted to
determine the cross-reactivity of non-related compounds with the test at
concentrations much higher than normally found in the urine of people
using or abusing them. No cross reactivity was detected with the
substances listed in Table II.
Table
I: Concentrations
of drug-related compounds showing positive response approximately
equivalent to the cut-off set for the test:
The following
Methamphetamine-related substances yield positive results for
Methamphetamine:
(+) 3,4-Methylenedioxymethamphetamine (MDMA, Ecstacy)
500
ng/ml
d-Amphetamine
50,000 ng/ml
(+) 3,4-Methylenedioxyamphetamine (MDA)
50,000 ng/ml
(+) Methamphetamine
50,000 ng/ml
D,
1-amphetamine
10,000 ng/ml
Pseudoephedrine
1,000 ng/ml
Ephedrine
25,000 ng/ml
The following Marijuana-related substances yield positive results for
Marijuana:
|
11-Nor-D
8-THC-9-COOH
50 ng/ml
11-Nor-D 9-THC-9-COOH
50 ng/ml
D-8-THC
1 mg/ml
|
D-9-THC
4 mg/ml
Cannabinol
10 mg/ml
11-hydroxy-D
9-THC
10 mg/ml
|
The following
Cocaine-related substances yield positive results for Cocaine:
Benzoylecgonine
300 ng/ml
Cocaine
300 ng/ml
Isoxsurpine
1,500 ng/ml
Table II:
Compounds tested and found not to cross-react with the test at a 10mg/ml
concentration in urine.
The following
compounds do not cross-react with Methamphetamine:
The
following compounds do not cross-react with Phencyclidine:
3.
Accuracy:
The
accuracy of the Multi-drug Multi-drug Test was first tested in urine
samples in-house and subsequently in a clinical trial of urine samples
submitted to a NIDA certified laboratory. In both cases the laboratories
used EMIT II as their screening procedure. All positive samples by either
screening method were confirmed by GC/MS. The data was combined and the
results are summarized as follows:
Marijuana (THC)
Syva
EMIT II Positive Syva
EMIT II Negative
Positive
297
4
Negative
4
259
When compared to EMIT II the relative sensitivity
between positive samples was 98.67%. The relative specificity between
negative samples was 98.48%. The concordance of the combined data with
respect to EMIT II was 98.58%.
Cocaine (BEG)
Syva EMIT II Positive
Syva EMIT II Negative
Positive
194
1
Negative
0
212
When compared to EMIT II the relative sensitivity between positive
samples was 100%. The relative specificity between negative samples was
99.55%. The concordance of the combined data with respect to EMIT II was
99.75%.
Methamphetamine (MET)
Syva EMIT II
Positive Syva
EMIT II Negative
Positive
179
14
Negative
0
262
When compared to EMIT II the relative sensitivity between positive
samples was 100%. The relative specificity between negative samples was
94.9%. The concordance of the combined data with respect to EMIT II was
96.92%.
When compared to EMIT II the relative sensitivity between
positive samples was 100%. The relative specificity between negative
samples was 100%. The concordance of the combined data with respect to
EMIT II was 100%.
Note: The above data for
Methadone reflects the results of in-house urine sample testing only.
Bibliography
General
1. Baselt,
R.C., Disposition of Toxic Drugs and Chemicals in Man, 2nd Ed., Biomedical Publ., Davis,
CA, p.488 (1982).
2.
Cody, J.T., Schwarzhoff, R., J. Anal. Toxicol., 17: 2630 (1993).
3.
Urine Testing for Drugs of Abuse, NIDA Research Monograph 73,
(1986).
4. Dasguspta,
A., Saldana, S., Kinnaman, G., Smith M., Johansen, K., Clin. Chem., 39(1):
104-108 (1993).
5. Department
of Health and Human Services, Fed. Regist., 53(69): 11970-11989 (1988),
(1989).
6.
FDA Guidence for Labeling Urine Drugs of Abuse Screening Testing,
Kshitij Mohan, 7/21/1987.
|